LT560B3-2 with Low oxygen pressure alarm 10.4' inch anesthesia unit with respiratory
Main body:High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application :adult, child
Gas source : O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter:O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply:25~75L/min
Low oxygen pressure alarm :There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer:
it has the function of automatic compensation based on pressure, temperature, and flow rate.
The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit:working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
Respiratory Bellows:bellows for adults, bellows for children
tidal volume range:0~1,500 ml
Ventilator
Display mode:High-definition TFT color LCD screen display
Ventilation mode :IPPV, SIPPV, SIMV, PEEP,MANUAL,SIGH, (IMV, VCV, PCV)
Ventilation parameters:
Tidal volume 50~1500 ml (20~1500 ml)
Rate 1~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure -1.0~2.0Kpa
PEEP 0~2.0Kpa
Pressure Range 1.0~6.0Kpa
SIGH 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform 0~50%
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow
-volume oop,pressure-volume loop)
Monitoring of oxygen concentration 21%~100%
Security Alarm System
Oxygen Concentration alarm:
upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm:
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm : it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm:it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power :AC 220V±10% 50Hz±1Hz(OptionalUPS, Storage battery)
(Mechanical Arms) (End-expiratory CO2 monitoring)
Items within brackets are optional.