LT2000B2 Hospital Mobile Medical Ventilator
Feature:
Display mode High-definition 8.4” TFT color LCD screen display
Ventilation mode IPPV, SIPPV, SIMV, PEEP,MANUAL,SIGH
Ventilation parameters Tidal volume 50~1500 ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure-1.0~2.0kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0kPa
SIGH 1.5 times the inspiratory time
Inspiratoryplatform 0~50%
Parameters for ventilation monitoring tidal volume,ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform
Monitoring of oxygen concentration 21%~100%
Security Alarm System
Oxygen Concentration alarm: upper limit setting range 50%~100%,low limit setting range 15%~50%
Airway pressure alarm :upper limit setting range 1.0~6.0kPa,low limit setting range 0.4~2.0kPa
Per-minute ventilation volume alarm : upper limit setting range 3.0~30L/min ,low limit setting range 1.0~10L/min
Sustained high-pressure alarm: it will give alarm when stress have consistently been higher than 2.5kPa
Suffocation alarm :it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power :AC 220V±10% 50Hz±1Hz
UPS, Storage battery
Mechanical Arms
End-expiratory CO2monitoring (choose and match)