1.Fluoroscopic Capacity Max rated capacity : Tube Current 4mA, Tube Voltage 120kV Automatic Fluoroscopy : Tube Voltage:40kV~120kV Adjust Automatically Tube Current : 0.3mA~4mA Manually Setting Manual Fluoroscopy : Continuous Tube Voltage:40kV~120kV Continuous Tube Current : 0.3mA~4mA Pulse Fluoroscopy : Continuous Tube Voltage 40kV~120kV Continuous Tube Current 4.1mA~30mA Pulse frequency : 0.1~25fps, Continuously Adjustable 2.Photography Capacity: Max rated capacity : 5 KW Tube Voltage and Current Combination : 40kV~49kV 1 ~180 mAs 50kV~59kV 1 ~140 mAs 60kV~69kV 1 ~125 mAs 70kV~79kV 1 ~110 mAs 80kV~89kV 1 ~100 mAs 90kV~99kV 1 ~80 mAs 100kV~109kV 1 ~63 mAs 110kV~120kV 1 ~50 mAs Plateholder Size : 200mm×250mm(8″×10″) or 250mm×300mm(10″×12″) 3.X-ray Tube X-ray Tube Special for High Frequency Fixed anode Dual-focus : small focus: 0.3 .Large focus:1.5mm Inverter Frequency : 40KHz Thermal capacity : 35KJ (47HU) 4.Video System : Image Intensifier : Image Intensifier made by TOSHIBA (9″, 6″,4.5″) CCD Video camera : 1.0 mega pixels ultra low-light CCD camera imported from Japan. Monitor : High-definition, high-resolution, optional specification 5.Image Processing System: (standard) : real-time image collection, specular image , recursive level 0-15 continuously adjustable,last frame frozen, Gamma correction, mass storage,image W/L adjustable, gray-scale conversion,interest district balance, reversal, noise reduction,intensify, smooth, sharpen, compress, enlarge,measurement, label, print and typesetting, DicomSCP, DicomSCU, DicomDIR, Dicom Printing,HIS 6.Structure Directive wheel : ±90°revolution,can freely change the moving direction of the unit. Ascending & Descending Range of Pillar : ≥400mm C-arm : Forward and Backward Movement: 200mm, Revolution around Horizontal Axis: ±180° Revolution around Vertical Axis : ±15°, easier to take special positions radiography ) Focus screen distance:1000mm, C-arm opening :800mm, C-arm arc depth : 640mm Slip on orbit : 120°(+90°~ -30°) |