Display mode:High-definition TFT color LCD screen display
Ventilation mode :IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV)
Ventilation parameters:
Tidal volume: 50~1500 ml (20~1500 ml)
Rate: 1~99 bpm
SIMV rate: 2~20 bpm
I:E: 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure: -1.0~2.0Kpa
PEEP: 0~2.0Kpa
Pressure Range: 1.0~6.0Kpa
SIGH: 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform: 0~50%
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow
-volume oop,pressure-volume loop)
Monitoring of oxygen concentration: 21%~100%
Security Alarm System
Oxygen Concentration alarm:
upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm:
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm : it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm:it will give sound and light alarm if there is no tidal volume input for 15 seconds