LTJH-2028A hotsale hospital hemodialysis machine for dialysis Centre | |
|
Clinical Application: | | | | | | | | | | | | |
| | | | | | | | | | | | | | | |
> Hemodialysis (HD) | | | | | | | | | | |
> Pure Ultrafiltration (ISO UF) | | | | | | | | | | |
> Hemoperfusion (HP) | | | | | | | | | | |
> Hemofiltration (HF) | | | | | | | | | | |
> Plasma Exchange (PE) | | | | | | | | | | |
| | | | | | | | | | | | | | | |
Product Features: | | | | | | | | | | | | |
| | | | | | | | | | | | | | | |
· LCD touch screen with high definition(10.4"). | | | | | | |
· Automation: Self-test, rinse, disinfection, hot disinfection, ISO UF, shutdown, etc. Automatically start treatment once blood flow sensed. | | | | | |
· Bicarbonate dry powder cartridge. | | | | | | |
· Use with dialysis fluid of various formulas (acetate & bicarbonate). | | | | | | |
· Kt/V function. | | | | | | |
· Double pumps: blood pump and spare pump providing more treatment options; Compatible with blood lines of different diameters. | | | | | | |
· Single & Double needle dialysis. | | | | | | |
· Using magnetic valves made in Switzerland. | | | | | | |
· In-built battery backup for max. 30 minutes. | | | | | | |
· Dialyzate flux linearity adjustable (300~800mL/min). | | | | | | |
· Support bedside treatment. | | | | | | |
· HCO3 adjustment for serious acid toxicosis patient. | | | | | | |
| | | | | | | | | | | | | | | |
| | | | | | | | | | | |
Safety Guaranteed: | | | | | | | |
| | | | | | | | | | | | | | | |
Audible & visible alarm with warning messages & troubleshooting. | | | | | |
· Blood flow monitoring. | | | | | | |
· Arterial and Venous pressure monitoring. | | | | | | |
· Transmembrane pressure(TMP) monitoring. | | | | | | |
· Non invasive blood pressure(NIBP) monitoring. | | | | | | |
· Air Monitoring: Air bubble detector using infrared sensor; Blood level detector using ultrasonic sensor. | | | | | |
· Blood leak monitoring. | |
· Water way entirely aparts from electric way. | | | | | | |
· Dialyzate temperature is selectable between 35 and 39 ℃. | | | | | | |
· Conductivity is controllable. | | | | | | |
· Sodium and ultrafiltration profiles for various treatment conditions. | | | | | | |
· Heparin pump timing for safety and convenience. | | | | | | |
· Disinfection: Rinsing, hot disinfection, chemical disinfection; Double-circulation disinfection system: Fluid supply loop and return loop are completely separated during disinfection to avoid pollution of the whole machine in case of mis-sucking disinfection liquid. This gives protection from intercross infection. | | | | | | |
| | | | | | | | | | | | | | Page 1/2 | |
Technical Parameters: | | | | | | | |
| | | | | | | | | | | | | | | |
General Data | Electrical Data |
· Dimensions: 370x340x1420mm (LxWxH) | · Power: 1500W · Frequency: 50/60Hz |
· Weight: 90 Kg | · Voltage: AC220V±10% (110V optional) |
· Water inlet pressure: 0~0.6MPa | Dialysis Fluid Circuit |
· Water inlet temperature: 5~30℃ | · Dialysate flow range: linearity adjustable 300~800mL/min |
· Ambient temperature: 10~30℃ | · Dialysate temperature range: Selectable 35~39℃ |
· Relative Humidity ≤70% | · Dialysate conductivity range: 13~15.5mS/cm (±0.1mS/cm) |
Extracorporeal Circuit | · UF flow range: 0~4000mL/hr (±30mL/hr) |
· Arterial pressure monitoring: | · ISO UF flow range: 0~2000mL/hr (±30mL/hr) |
Display range: -300~+300mmHg (±10mmHg) | · Blood leak detector: |
· Venous pressure monitoring: | Automatically close the blood loop when blood leak ≥1mL per liter dialysate (Flow: 500mL/min) |
Display range: -50~+300mmHg (±10mmHg) |
· TMP range: -100~+600mmHg (±20mmHg) | Disinfection and Cleaning |
· Blood pump/ Spare pump: | · Disinfection: Use citric acid, peracetic acid and sodium hypochlorite. |
Flow range: 15~340mL/min (φ 6mm) 20~620mL/min (φ 8mm) | · Hot rinse temperature: 90℃ |
| | | | | | | |
· Blood level detector: ultrasonic sensor. | | | | | | | | |
· Air bubble detector: infrared sensor. | | | | | | | | |
Threshold of reaction: single air bubble of 200 μI exists when blood flux is 200mL/min | | |
| | | | | | | |
· Heparin pump: | | |
Delivery range: 0.1~10mL/hr (±5%) Accuracy: 0.1mL/hr Injector size: 10/20/30mL, 20/30/50mL (optional) | | |
| |
· Warranty: 18 months from the date of delivery |
· After-sale Service: Free spare parts will be provided for necessary replacements within warranty period; After warranty period, spare parts will be provided at cost price. All parts will be available for at least 10 years. |
· Packaging: Wood case; Dimensions: 171x67x78cm; N.W./G.W.: 110kg/130kg |