1.Main body:High-strength engineering plastic rack, light, beautiful and corrosion resistant
2.Scope of application :adult, child
3.Gas source :O2: 0.27~0.55MPa
4.Flow meter :O2: 0.1~2L/min 2.5~10L/min
5.Flow rate of rapid oxygen supply :25~75L/min
6.Low oxygen pressure alarm:There will be sound alarm when the oxygen pressure < 0.2MPa
7.Vaporizer
it has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
8.Respiratory circuit
working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
9.Respiratory bellows :bellows for adults, bellows for children tidal volume range:0~1500 ml
Ventilator:
10.Display mode :High-definition 5.7"LCD screen display
11.Ventilation mode:IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
12.Ventilation parameters :
Tidal volume:50~1500ml
Rate:2~99 bpm
SIMV rate :2~20 bpm
I:E:2:1~1:8
Inspiratory trigger pressure : -1.0~2.0 kPa
Pressure range : 1.0~6.0 kPa
13.Parameters for ventilation monitoring:tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
14.Security Alarm System
Airway pressure alarm
upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm :it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm:it will give sound and light alarm if there is no tidal volume input in 15s
Power alarm
15.Power :AC 220V±10% 50Hz±1Hz